Function Description – MI273 through MI297 – Signed Integer
Bit 15 - Clear
NAME BITS FUNCTION DESCRIPTION
Fault Relay Active 0 Set if True
Sensor Missing – Sensor 1 1 Set if True
Sensor Missing – Sensor 2 2 Set if True
Sensor Missing – Sensor 3 3 Set if True
Calibration Fault – Sensor 1 4 Set if True
Calibration Fault – Sensor 2 5 Set if True
Calibration Fault – Sensor 3 6 Set if True
Power Fail Fault – Sensor 1 7 Set if True
Power Fail Fault – Sensor 2 8 Set if True
Power Fail Fault – Sensor 3 9 Set if True
Power Fault – Main Unit +5VDC 10 Set if True
Sensor End of Life – Sensor 1 11 Set if True
Sensor End of Life – Sensor 2 12 Set if True
Sensor End of Life – Sensor 3 13 Set if True
Configuration Reset 14 Set if True
Sign 15 Always 0
Bit 15 - Set
NAME BITS FUNCTION DESCRIPTION
LOC – Sensor 1 0 Set if True
LOC – Sensor 2 1 Set if True
LOC – Sensor 3 2 Set if True
Over Range – Sensor 1 3 Set if True
Over Range – Sensor 2 4 Set if True
Over Range – Sensor 3 5 Set if True
Quick Under Range – Sensor 1 6 Set if True
Quick Under Range – Sensor 2 7 Set if True
Quick Under Range – Sensor 3 8 Set if True
Under Range – Sensor 1 9 Set if True
Under Range – Sensor 2 10 Set if True
Under Range – Sensor 3 11 Set if True
Not Used 12 Not Used
Not Used 13 Not Used
Not Used 14 Not Used
Sign 15 Always 1
B-8