Operation
CONFIGURE
EDIT NAME Enter a 10-digit name in any combination of symbols and alpha or numeric characters.
SELECT MEASURE Select the appropriate measurement units to display.
DISPLAY FORMAT Select the measurement resolution (xx.xx pH or xx.x pH).
TEMP UNITS Choose from the displayed options (°C or °F).
LOG SETUP
Choose SENSOR INTERVAL to set the sensor log interval or select TEMP INTERVAL to set the
temperature log interval.
REJECT FREQUENCY
Choose 50 or 60 Hz depending on the power line frequency for optimal noise rejection. Default is
60 Hz.
FILTER Select 0–60 second signal averaging time.
TEMP ELEMENT Select type of temperature element from the displayed choices.
SELECT BUFFER Select the buffer type (standard 4, 7, 10 or DIN 19267) from the displayed choices.
PURE H20 COMP
Allows the user to specify that ammonia, morpholine, or other user-defined electrolyte is being
used in the application, allowing a temperature-dependent linear slope factor to be applied to the
measured pH.
CAL DAYS Number of days since the last calibration. Default notification at 60 days.
SENSOR DAYS Number of days the sensor has been in operation. Default notification at 365 days.
DEFAULT SETUP Resets all user-editable options to their factory-defaults.
DIAG/TEST
PROBE INFO
Display the sensor type, entered name of the sensor (Default: sensor serial number.), the sensor
serial number, the software version number, and the sensor driver version number.
CAL DATA Displays the pH slope and the date of the last calibration
SIGNAL
SENSOR SIGNAL: Displays the sensor output in mV
SENSOR ADC COUNTS: Displays the sensor ADC counts
TEMP ADC COUNTS: Displays raw data for temperature ADC counts. ADC counts are
comparable to A/D counts and are for sensor electronic diagnostic purposes only.
ELECTRODE STATE: Identifies the state of the electrode (good or bad) depending on whether
the impedance is within preset limits.
ACTIVE ELECT: Displays the impedance (Mohms) of the active electrode if Imped Status is set to
Enabled.
REF. ELECTRODE: Displays the impedance (Mohms) of the reference electrode if Imped Status
is set to Enabled.
IMPED STATUS: Sensor diagnostic. Choose Enabled or Disabled.
COUNTERS
SENSOR DAYS: displays the cumulative days the sensor has been in use.
RESET SENSOR: Allows the sensor counter to be reset to zero.
ELECTRODE DAYS: Cumulative days the electrode has been in use.
5.4 pH Sensor Setup Menu (continued)