In ScanWorld “
Station-004:SW-004
” two targets were acquired incorrectly. Target 25 was fitted
incorrectly and requires re-fitting. Target 33 is positioned correctly, but was named wrong. It
requires relabeling to TargetID 3.
Delete Targets in ControlSpaces
ControlSpaces contain constraint objects, but you cannot create any objects in ControlSpaces.
They are always created in either a ModelSpace or TruSpace and then copied across to selective
ControlSpaces. We will therefore first delete the incorrect targets in the ControlSpaces and then
repopulate them with the corrected targets created in the ModelSpace.
Figure 15
•
Open
ControlSpace
under
Station-004:SW-
004
•
Navigate to the target 25.
•
See that the vertex is not accurately placed
in the center of the target (
source not found.
).
•
Select
the vertex and
delete
.
•
Confirm the upcoming message about
deleting objects from a
ControlSpace.
•
Now navigate to target 33.
•
Delete
also this vertex.
•
Close
the ControlSpace.
•
Repeat the same process also in
ControlSpace [Registration1]
.
ControlSpaces are not directly linked so that they can contain different
constraints for different registrations. Therefore the clean-up has to be done in
in all affected ControlSpaces.
An alternative way to quickly delete all targets in a ModelSpace View is via the
View Properties Window (View | Set Object Visibility or Shift S).
Highlight „Vertex“ and press Select. Then delete all selected.
Recreate Target in ModelSpace
The next step is now to correctly re-create the target 25 and copy it across to all ControlSpaces.
•
In Cyclone Navigator navigate to
Station-004: SW-004
•
Open the ModelSpace 1 View 1
•
Delete the incorrectly placed vertex with TargetID 25
Target 25 was placed on a park bench far out from the scanner. A target fine-scan with a denser
spacing than the main scan has been executed in the field to be able to accurately extract this