GUI Reference
R&S
®
ZNB/ZNBT
528User Manual 1173.9163.02 ─ 55
Modify Frequency Conversion Dialog
The "Modify Frequency Conversion" dialog defines a port-specific frequency formula
(for frequency sweeps) or CW frequency (for power, time and CW Mode sweeps).
Access: Port Settings Dialog > "Arb Frequency" tab > "Freq. Conversion" cells
Dependencies
The analyzer firmware enforces the following rules:
For balanced ports, the constituent physical ports must use the same Frequency
Conversion Formula.
Gen ports driven by the same internal source (R&S ZNBT only) must use the same
frequency conversion formula.
Current Formula
Apart from the current Frequency Conversion Formula, the analyzer firmware main-
tains a list of named, user-defined formulas.
"Add Formula to List..." adds the current frequency conversion formula to this list.
"Remove Selected Formula" removes the selected user-defined formula from this
list.
"Change Formula Name..." allows you to change the name of the currently
selected user-defined formula.
Note:
User-defined formulas are persisted in the recall set. A preset deletes all user-
defined formulas.