GUI Reference
R&S
®
ZNB/ZNBT
478User Manual 1173.9163.02 ─ 55
Remote command:
[SENSe<Ch>:]CORRection:CONNection
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe<Ch>:]CORRection:CONNection:DELete
6.11.2.2 Calibration Kits Dialog
The "Calibration Kits" dialog shows the available calibration kits for the different con-
nector types. It is also used for cal kit and cal kit file management.
Access: CHANNEL – [CAL] > "Cal Devices" > "Cal Kits..."
Related information
Refer to the following sections:
●
See also Chapter 5.5.2, "Calibration Standards and Calibration Kits", on page 159
●
Chapter 5.5.2.4, "Cal Kit Files", on page 164
●
Chapter 5.5.2.2, "Cal Kit Parameter Types", on page 162
●
Chapter 6.11.2.1, "Cal Connector Types Dialog", on page 476
The contents of the "Available Cal Kits" table vary, depending on the selected "Connec-
tor Type". The table can also contain kits with ideal or typical parameter values; see
Cal Kit Parameter Types. The "Agilent Model" is an optional scheme to characterize
the offset parameters of the standards; see "Offset Parameters" on page 482.
Cal kit labels
Assigning a "Label" to user-defined calibration kits is optional. However, the label is
displayed in many dialogs and can provide useful information about the kit, e.g. its
serial number. It is even possible to assign several calibration kits with the same name,
distinguished by their label, to a common connector type. See also Chapter 8.3.14.6,
"[SENSe:]CORRection:CKIT... with Labels", on page 1125.
Cal Softtool