EasyManua.ls Logo

R&S ZNB - 6.11.3.1 Controls on the Pwr Cal Settings Tab

R&S ZNB
1496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI Reference
R&S
®
ZNB/ZNBT
488User Manual 1173.9163.02 ─ 55
Efficient power calibration procedure
For standard applications, open the "Start Cal" tab and select "Scalar Power Cal" –
"Power Cal..." to perform the necessary calibration sweeps with default power cali-
bration settings. You do not need any of the buttons in the "Pwr Cal Settings" tab.
Select "Cal Power..." if you use an amplifier between the source port and the DUT.
Select "Transm. Coefficients..." if you want to modify the power calibration proce-
dure.
Background information
Refer to Chapter 5.5.6, "Scalar Power Calibration", on page 173.
6.11.3.1 Controls on the Pwr Cal Settings Tab
Some buttons serve as openers for additional dialogs:
"Cal Power...": see Chapter 6.11.3.2, "Modify Cal Power Dialog", on page 490
"Transm. Coefficients...": see Chapter 6.11.3.3, "Power Meter Transmission Coeffi-
cients Dialog", on page 492
"Power Meters...": see Chapter 5.7.25, "External Power Meters", on page 241
Switch Off Other Sources
Ensures that the power at all sources except the calibrated source is switched off dur-
ing the calibration. Recommended, if the measurement involves a combination of dif-
ferent signals.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:OSOurces[:STATe]
Cal Softtool

Table of Contents

Related product manuals