EasyManua.ls Logo

R&S ZNB - 6.11.1.2 Channel Selection for Calibration Dialog; 6.11.1.3 Calibration Unit Wizard

R&S ZNB
1496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI Reference
R&S
®
ZNB/ZNBT
415User Manual 1173.9163.02 ─ 55
If the active recall set contains several channels, an "Info" box lets you decide whether
the R&S ZNB/ZNBT acquires calibration data for the active channel or for all channels.
SMARTerCal – Start... (Cal Unit)
Depending on the Multiple Cal in Calibration Wizard setting, the "Start... (Cal Unit)" but-
ton either opens the SMARTerCal Wizard (Cal Unit) or the SMARTerCal Wizard (Cal
Unit, MultiCal).
SMARTerCal – Start... (Manual)
Depending on the Multiple Cal in Calibration Wizard setting, the "Start... (Cal Unit)" but-
ton either opens the SMARTerCal Wizard (Manual) or the SMARTerCal Wizard (Man-
ual, MultiCal) .
SMARTerCal – Repeat...
Reenters the data acquisition step of the current SMARTerCal, restoring the related
channel settings. This function provides a convenient way to repeat or correct an exist-
ing SMARTerCal without necessarily repeating all measurements.
6.11.1.2 Channel Selection for Calibration Dialog
If you initiate a calibration and there is more than one channel in the current recall set,
you are prompted to select the channels to be calibrated:
Use
"Active Channel" if you only want to calibrate the active channel
"All Channels" if you want to calibrate all channels in the current recall set
"Selected Channels" if you want to calibrate only certain channels
6.11.1.3 Calibration Unit Wizard
The "Calibration Unit" wizard guides you through the setup and execution of an auto-
matic calibration.
Access: CHANNEL – [CAL] > "Start Cal" > Calibration – "Start... (Cal Unit)"
Cal Softtool

Table of Contents

Related product manuals