EasyManua.ls Logo

R&S ZNB - 5.5.7 SMARTerCal

R&S ZNB
1496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Concepts and Features
R&S
®
ZNB/ZNBT
179User Manual 1173.9163.02 ─ 55
B: Two-port at power meter (during calibration)
Test and measurement procedure:
1. Perform the calibration with the additional two-port between the analyzer port and
the power sensor. During the calibration, the analyzer increases the power sensor
values by the 2-port transmission coefficients to move the calibration plane of the
power calibration towards the input of the DUT. The calibration plane corresponds
to the input of the additional 2-port.
2. Perform the measurement without the additional two-port.
Practical example: An adapter or attenuator with known attenuation is needed to con-
nect the power sensor to the test port of the network analyzer. The transmission coeffi-
cients of the adapter are used for the power meter correction.
5.5.7 SMARTerCal
A SMARTerCal (smarter calibration) is a combination of a full n-port system error cor-
rection (TOSM, UOSM, Adapter Removal, TRL, TNA ...) for two or more ports with a
(scalar) receiver power calibration at a single port. The two calibration methods serve
different purposes:
The system error correction requires a set of calibration standards; it provides vec-
tor error-corrected S-parameters. For equal port frequencies, the n-port calibration
types provide the full set of error terms. For frequency conversion measurements,
a source match correction and (optional) load match correction is calculated.
The receiver power calibration requires an external power meter; it corrects the
power readings of the reference and measurement receivers according to the mea-
sured absolute power at the calibration plane. This does not include a readjust-
ment of the actual source power (flatness calibration).
Example: Channel base power: 10 dBm; the test setup involves a 3-dB attenua-
tion between the source port and the calibration plane. After the power calibration
is applied, the analyzer indicates an output power (a-wave) of 13 dBm, although
the actual source power remains at 10 dBm.
The SMARTerCal is also applied to ratios and wave quantities. For measurements on
linear DUTs, SMARTerCal is sufficient. Non-linear measurements can be further
improved by a combination of a SMARTerCal plus a scalar power calibration. See
Combining SMARTerCal with Scalar Power Calibration.
Calibration

Table of Contents

Related product manuals