Command Reference
R&S
®
ZNB/ZNBT
1118User Manual 1173.9163.02 ─ 55
8.3.14.5 [SENSe:]CORRection:CKIT...
The [SENSe:]CORRection:CKIT... commands deal with calibration kits and cal kit
data. The calibration kits are distinguished by their names (<CalkitName>), the optional
labels (<label>) can be used to carry information about the calibration standard.
In order to handle several identical calibration kits with different serial numbers use the
commands of Chapter 8.3.14.6, "[SENSe:]CORRection:CKIT... with Labels",
on page 1125.
[SENSe:]CORRection:CKIT:CATalog?.............................................................................1118
[SENSe:]CORRection:CKIT:DELete................................................................................1118
[SENSe:]CORRection:CKIT:DMODe...............................................................................1119
[SENSe:]CORRection:CKIT:LABel..................................................................................1119
[SENSe:]CORRection:CKIT:SELect................................................................................1120
[SENSe:]CORRection:CKIT:STANdard:CATalog?.............................................................1120
[SENSe:]CORRection:CKIT:STANdard:DATA?................................................................ 1120
[SENSe:]CORRection:CKIT:<ConnType>:SELect............................................................ 1121
[SENSe:]CORRection:CKIT:<StandardType>.................................................................. 1122
[SENSe:]CORRection:CKIT:CATalog? [<ConnectorType>]
Returns a list of all cal kits for a given connector type or for all connector types.
Query parameters:
<ConnectorType> Name of the connector type (optional). Use [SENSe<Ch>:
]CORRection:CONNection:CATalog? to query connector
names.
If omitted, the command returns the list of all cal kits.
Example: See [SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage: Query only
Manual operation: See "Available Cal Kits" on page 479
[SENSe:]CORRection:CKIT:DELete <CalKitName>
Deletes an imported or user-defined cal kit.
Note: It is not possible to modify or store predefined or ideal kits.
Setting parameters:
<CalKitName> String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
Example: See [SENSe:]CORRection:CKIT:LABel
Usage: Setting only
Manual operation: See "Add / Copy / Delete / Standards..." on page 479