GUI Reference
R&S
®
ZNB/ZNBT
323User Manual 1173.9163.02 ─ 55
A "Notch" filter rejects all information in the specified time region and passes every-
thing else.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE]
Shape
Selects a gate shape which the R&S ZNB/ZNBT uses to filter the trace in the time
domain. The drop-down list visualizes how the time gate will affect a constant function
after transformation back into the frequency domain. The selected window is applied to
the active trace. The two red vertical lines represent the "Start" and "Stop" values
defining the size of the time gate.
See also Chapter 5.7.2.5, "Time Gates", on page 206.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
Side Lobe Level
Defines the side lobe suppression for an "Arbitrary Gate Shape (Dolph-Chebychev)"
gate. The entered value is the ratio of the power of the central lobe to the power of the
first side lobe in dB.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev
6.5.7 Distance to Fault Tab
This tab is only visible, if option K3 is installed on the R&S ZNB/ZNBT. Replicates the
"Distance to Fault" tab of the "Distance to Fault" application. See Chapter 6.15.4, "Dis-
tance to Fault Application", on page 674.
6.5.8 Trace Statistics Tab
Evaluates statistical and phase information of the entire trace or of a specific evalua-
tion range and calculates the x-dB compression point.