GUI Reference
R&S
®
ZNB/ZNBT
288User Manual 1173.9163.02 ─ 55
Remote command:
[SENSe<Ch>:]PAE:TYPE
Formula
Shows the formula for the calculation of the DC supply power P
DC
, depending on the
selected Measurement Type. The parameters R, U
0
, and/or I
0
must be entered in
accordance with the properties with the DC power source and precision resistor in use.
Remote command:
[SENSe<Ch>:]PAE:PARameters:I
[SENSe<Ch>:]PAE:PARameters:R
[SENSe<Ch>:]PAE:PARameters:U
In/Out
Define the logical ports for the RF input and output signal, respectively.
Remote command:
CALCulate<Ch>:PARameter:MEASure 'TraceName', 'PAE21 | ...'
CALCulate<Ch>:PARameter:SDEFine 'TraceName', 'PAE21 | ...'
U
m
/U
m1
/U
m2
/
Select the DC INPUT connectors used to determine the DC supply power.
How many of them are required depends on the selected Measurement Type.
Remote command:
[SENSe<Ch>:]PAE:DCINput:MAIN
[SENSe<Ch>:]PAE:DCINput:SECondary
Range DC
Select the DC range of each DC INPUT connector in use according to the (estimated)
input voltage range; see "DC 1/DC 2/DC 3/DC 4/Source Port" on page 286.
Remote command:
[SENSe<Ch>:]DC<DCInp>:RANGe
6.3 Format Softtool
The "Format" softtool allows you to define how the measured data is presented in the
diagram area.
Access: TRACE – [FORMAT]
Format Softtool