GUI Reference
R&S
®
ZNB/ZNBT
669User Manual 1173.9163.02 ─ 55
Depending on the "Measurement Selection" on page 663 either one, two, or three
channels must be set up:
one channel for S-parameter measurements
one channel for intermodulation measurements
Summary
The "Summary" page of the "DUT Sentric Wizard"
summarizes which channels will be created and which parameters they will mea-
sure
allows you where to define these channels (new setup, append to current setup,
replace current setup)
allows you to decide whether to proceed with a (suitable) calibration
Calibrate Newly Created Channels
If you select "Calibrate Newly Created Channels", then on closing the "DUT Centric
Wizard", the FW will create the channels and proceed with suitable calibrations.
Guided Calibration Wizard
The "Guided Calibration" wizard lets you calibrate a set of channels using the calibra-
tion methods the firmware proposes for the related measurements.
If a calibration unit is connected, it is automatically used.
Access: CHANNEL – [CAL] > "Start Cal" > "Start... (Guided)"