GUI Reference
R&S
®
ZNB/ZNBT
480User Manual 1173.9163.02 ─ 55
[SENSe:]CORRection:CKIT:DMODe
[SENSe:]CORRection:CKIT:DELete
Query connector types and calibration kits:
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe:]CORRection:CKIT:CATalog?
Import Cal Kit... / Export Cal Kit...
The buttons below the "Connector Type" list are used to store cal kit data to a file and
to reload previously stored cal kit files. By default, calibration kit files are stored in the
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration directory;
see Chapter 5.5.2.4, "Cal Kit Files", on page 164.
Remote command:
MMEMory:LOAD:CKIT
MMEMory:STORe:CKIT
Kit Standards Dialog
The "Kit Standards" dialog shows the calibration standards in a selected calibration kit.
It is also used to modify the contents of a user-defined kit.
Related information
Refer to the following sections:
●
Chapter 5.5.2.4, "Cal Kit Files", on page 164
●
Chapter 6.11.2.2, "Calibration Kits Dialog", on page 478
●
Chapter 5.5.2.1, "Calibration Standard Types", on page 160
Access: Calibration Kits Dialog > "Standards..."
Cal Softtool