Command Reference
R&S
®
ZNB/ZNBT
1019User Manual 1173.9163.02 ─ 55
Manual operation: See "RFFE columns (sweep sequencer table)" on page 522
CONTrol<Ch>:SEQuence:CLEar:ALL
For unsegmented sweeps this command deletes the command/switch sequence for
the respective channel (defined using CONTrol<Ch>:SEQuence<Nr>:RFFE<Bus>:
COMMand:DATA and CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:VOLTage).
Suffix:
<Ch>
.
Channel number
Usage: Event
Manual operation: See "Range" on page 522
CONTrol<Ch>:SEQuence:COUNt?
For unsegmented sweeps this command queries the length of the command sequence
defined for a particular channel (using CONTrol<Ch>:SEQuence<Nr>:RFFE<Bus>:
COMMand:DATA).
Suffix:
<Ch>
.
Channel number
Usage: Query only
Manual operation: See "Range" on page 522
CONTrol<Ch>:SEQuence<Nr>:DELay <Delay>
For unsegmented sweeps this command allows to introduce delays between subse-
quent sequence steps and between the final sequence step and the sweep (segment)
start.
The command/switch sequences are defined using CONTrol<Ch>:SEQuence<Nr>:
RFFE<Bus>:COMMand:DATA and CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:
VOLTage.
Suffix:
<Ch>
.
Channel number
<Nr> Sequence number, defining the order in which the commands
shall be executed.
For every channel and RFFE interface, the sequence numbers
must be consecutive, starting at 1; reusing a sequence number
overwrites a previously defined command.
Parameters:
<Delay> Delay time
Default unit: s
Manual operation: See "Wait (Sweep Sequencer Table)" on page 523
SCPI Command Reference