[SENSe:]CORRection:COLLect:AVERage <Average>
Activates automatic averaging, which means that the VNA may perform multiple cali-
bration sweeps and apply averaging to reduce trace noise. In contrast to manual aver-
aging (see [SENSe<Ch>:]AVERage[:STATe]) the number of calibration sweeps is
calculated automatically.
Parameters:
<Average> AUTO | MANual
Manual operation: See "Auto Averaging" on page 701
[SENSe:]CORRection:COLLect:CHANnels:ALL <Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
●
Manual SMARterCal
●
Automatic SMARTerCal (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean> ON – calibrate all channels.
OFF – calibrate the active channel only.
Example: See [SENSe<Ch>:]CORRection:COLLect:SAVE:
SELected[:DUMMy]
Manual operation: See "Calibrate all Channels" on page 431
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes <Boolean>
Toggles the Multiple Calibration TYpes mode that allows you to calibrate a subset of
the available channels using channel-specific ports and calibration types (see Chap-
ter 5.5.8, "Parallel Calibration of Multiple Channels", on page 181).
Enable the MCTYpes mode before defining the calibrations to be performed using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine. Disable the MCTYpes
mode after the calibrations were saved using [SENSe<Ch>:]CORRection:
COLLect:SAVE:SELected[:DUMMy]. The latter must be called only once, with <Ch>
referring to one of the channels to be calibrated (see the example below).
Parameters:
<Boolean>