Command Reference
R&S
®
ZNB/ZNBT
1141User Manual 1173.9163.02 ─ 55
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe] <Boolean>
Selects an Auto Length (and Loss) calculation or a Direct Compensation.
Parameters:
<Boolean> 1 - Auto Length (and Loss), depending on the last
[SENSe<Ch>:]CORRection:COLLect:FIXTure:
LMParameter:LOSS[:STATe] setting
0 - Direct Compensation
*RST: 1
Example: See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:
ACQuire]
Manual operation: See "Direct Compensation" on page 583
[SENSe:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe] <Boolean>
Selects an Auto Length or an Auto Length and Loss calculation.
Parameters:
<Boolean> 1 - Auto Length and Loss
0 - Auto Length, no loss
*RST: 1
Example: See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:
ACQuire]
Manual operation: See "Auto Length / Auto Length and Loss" on page 582
[SENSe:]CORRection:COLLect:PMETer:ID <PowerMeter>, <TestPort>
Selects an external power meter for the SMARTerCal and assigns it to an analyzer
port.
Note: The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Parameters:
<PowerMeter> Number of external power meter. The parameters UP, DOWN,
MIN, MAX are not available for this command.
Range: 1 to number of configured external generators
*RST: The power meter selection is not changed by a
reset of the analyzer.
<TestPort> Test port number. During power calibration, the power meter is
connected to this port.
*RST: 1 to the number of test ports
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
Manual operation: See "Power Meter" on page 452
SCPI Command Reference