EasyManua.ls Logo

R&S ZNB - Page 1183

R&S ZNB
1496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1183User Manual 1173.9163.02 ─ 55
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
<TestPort1>, <TestPort2>...
Similar logic as [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:
DEFine:DEFault, but initially defines the default port assignments only by their
underlying test port set(s); the connected calibration unit ports are auto-detected at
the start of the corresponding calibration sweep.
The number of created assignments can be queried using the [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:COUNt? command.
See Chapter 5.5.5.4, "Multiple Port Assignments", on page 171 for background infor-
mation.
Suffix:
<Ch>
.
Setting parameters:
<TestPort1>
<TestPort2>
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1184
Usage: Setting only
Manual operation: See "Default Port Assignment" on page 421
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DELete:ALL
Deletes all port assignments (see Chapter 5.5.5.4, "Multiple Port Assignments",
on page 171).
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine
Usage: Event
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
Starts the automatic calibration sweep for the indicated channel and port assignment.
A complete, valid set of port assignments must be defined before you can initiate a cal-
ibration. See Chapter 5.5.5.4, "Multiple Port Assignments", on page 171 for back-
ground information.
Suffix:
<Ch>
.
Number of the calibrated channel
<Asg> Number of the port assignment
SCPI Command Reference

Table of Contents

Related product manuals