Command Reference
R&S
®
ZNB/ZNBT
1224User Manual 1173.9163.02 ─ 55
Parameters:
<StartFreq> Start frequency of the segment; see [SENSe<Ch>:
]SEGMent<Seg>:FREQuency:STARt.
Default unit: Hz
<StopFreq> Stop frequency of the segment; see [SENSe<Ch>:
]SEGMent<Seg>:FREQuency:STOP.
Default unit: Hz
<Points> Number of sweep points in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:SWEep:POINts.
<Power> Internal source power in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:POWer[:LEVel].
Default unit: dBm
<SegmentTime> Duration of the sweep in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:SWEep:TIME. In the setting [SENSe<Ch>:
]SEGMent<Seg>:INSert:SELect DWELl, this parameter is
replaced by <MeasDelay>.
Range: Depending on other channel settings. AUTO acti-
vates automatic sweep time setting in the segment,
which is equivalent to the minimum sweep time
possible.
Default unit: s
<MeasDelay> Delay for each partial measurement in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect SWTime,
this parameter is replaced by <SegmentTime>.
<Unused> Ignored parameter (for compatibility with R&S ZVR analyzers).
Should be set to the default value 0.
<MeasBandwidth> IF bandwidth in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:BWIDth[:RESolution].
Default unit: Hz
<LO> POSitive | NEGative | AUTO
Position of the local oscillator frequency LO relative to the RF
frequency. In remote control this parameter must be set when a
sweep segment is created.
See [SENSe<Ch>:]FREQuency:SBANd.
<Selectivity> NORMal | MEDium | HIGH
Selectivity of the IF filter. See [SENSe<Ch>:]SEGMent<Seg>:
BWIDth[:RESolution]:SELect.
<Mode> AUTO
Ignored.
SCPI Command Reference