Command Reference
R&S
®
ZNB/ZNBT
1276User Manual 1173.9163.02 ─ 55
<PhyPt> Calibrated port number. This parameter is ignored; the tolerance
value is valid for all sources.
Parameters:
<Tolerance> Tolerance value
Range: 0.001 dB to 1000 dB
*RST: 1 dB
Default unit: dB
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Tolerance" on page 489
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:CFACtor
<Convergence>
Specifies the convergence factor for a source power calibration.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the con-
vergence factor is valid for all sources.
<PhyPt> Calibrated port number. This parameter is ignored; the conver-
gence factor is valid for all sources.
Parameters:
<Convergence> Convergence factor
Range: 0 to 2
Increment: 0.1
*RST: 0.8
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Convergence" on page 489
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:FLATness <arg0>
Enables or disables the source power calibration. The command is valid for all chan-
nels, ports and external generators.
It is not possible to disable flatness calibration and the reference receiver calibration
simultaneously
(SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:RRECeiver).
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the verifi-
cation is either enabled or disabled for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the verifica-
tion is either enabled or disabled for all sources.
SCPI Command Reference