EasyManua.ls Logo

R&S ZNB - Page 1302

R&S ZNB
1496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1302User Manual 1173.9163.02 ─ 55
SYSTem:COMMunicate:AKAL:CONNection <CalStandard>, <Port>[, <SecondPort>,
<CalkitIndex>]
Connects the selected calibration standard to one or two ports of the active calibration
unit (see SYSTem:COMMunicate:RDEVice:AKAL:ADDRess).
Setting parameters:
<CalStandard> THRough | OPEN | SHORt | MATCh
Connected one or two-port standard
<Port> Port number of the calibration unit, for one and two-port stand-
ards.
<SecondPort> Second port number of the calibration unit.
For two-port standards (THRough) this parameter must be provi-
ded, for one-port standards (OPEN | SHORt | MATCh) it must be
omitted.
<CalkitIndex> For calibration units that are equipped with more than one cali-
bration kit, this parameter indicates the 1-based index of the cal
kit to be used (mandatory!).
For other calibration units, this parameter is optional, but if it is
provided, it must be set to 1.
Example:
SYST:COMM:AKAL:CONN THR, 1, 2
Connect a through standard between ports 1 and 2 of the cal
unit.
Usage: Setting only
Manual operation: See "Apply/Cancel" on page 424
SYSTem:COMMunicate:AKAL:MMEMory[:STATe] <Boolean>
Shows or hides the memory of the active calibration unit (see SYSTem:
COMMunicate:RDEVice:AKAL:ADDRess).
Setting parameters:
<Boolean> ON - memory is shown in a separate drive.
OFF - memory is not shown.
*RST: OFF
Example:
SYST:COMM:AKAL:MMEM ON
Show the memory of the active calibration unit.
Usage: Setting only
Manual operation: See "Apply/Cancel" on page 424
SYSTem:COMMunicate:CODec <Codec>
Selects the character encoding used at the remote interface. The selected encoding
applies to directory and file names, CalKit names, CalUnit characterizations and dis-
play titles.
SCPI Command Reference

Table of Contents

Related product manuals