Command Reference
R&S
®
ZNB/ZNBT
923User Manual 1173.9163.02 ─ 55
Parameters:
<Level> Reference level
*RST: 1 dB
Default unit: dB
Manual operation: See "Defined Value" on page 328
CALCulate<Chn>:STATistics:NLINear:COMP:RMARker <Marker>
Allows you to select the marker whose value shall be used as the reference ("small sig-
nal value") for the compression point calculation.
Only applies if CALCulate<Chn>:STATistics:NLINear:COMP:REFerence is set
to MARKer.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Marker> Marker number
If marker <Marker> is OFF for the related trace, the set com-
mand is silently ignored.
Manual operation: See "Selected Marker" on page 328
CALCulate<Chn>:STATistics:NLINear:COMP:RPHase <Phase>
Manually defines the reference phase for compression point calculation on phase for-
matted traces.
Only applies if CALCulate<Chn>:STATistics:NLINear:COMP:REFerence is set
to DVALue.
For dB formatted traces, the reference level can be set using CALCulate<Chn>:
STATistics:NLINear:COMP:RLEVel.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Phase> Reference phase
*RST: 1°
Default unit: deg
Manual operation: See "Defined Value" on page 328
CALCulate<Chn>:STATistics:NLINear:COMP:RPHase <Phase>
Manually defines the reference phase for compression point calculation on phase for-
matted traces.
Only applies if CALCulate<Chn>:STATistics:NLINear:COMP:REFerence is set
to DVALue.
SCPI Command Reference