Monitoringbasics
●Alarmsignalsareworkingandcanbeseenandheardinyourcareenvironment.
●Requiredparametercalibrationsarecompleted.
Supplymainsinterruption
Ifthesupplymainstotheequipmentisinterruptedforlessthan15minutes(andthe
B650isusedwithouttheoptionalbattery),themonitorkeepsthetrenddataand
thelatestuser-madesettings.Ifnot,contactauthorizedservicepersonnel.After15
minutes,allpatientinformationandtrenddataislostandthemonitorreturnstothe
userdefaultsettings(start-upmode).Formoreinformation,seethetechnicalmanual.
WARNING
TheB450mustalwaysbeusedwithabatteryinserted.This
willensurethefunctioningofthemonitorandpreventdata
lossduringpossiblesupplymainsinterruptions.
Powerfailurealarm
B650only.
Ifthereisapowerfailure(e.g.,thesupplymainsisinterruptedandthemonitorbattery
fails),themonitorgivesacontinuousbeepingalarm.Thisalarmremainsactiveforas
longasthereissomeresidualpowerleft,oruntilitissilencedwiththepauseaudio
alarmkey,oruntilthepowerisreconnected.
106CARESCAPEModularMonitors2062971-001