PatientSpirometry
2.SelectLoops.
3.SelectPrintLoop.
NOTE
Withcompactairwaymodules,youcanalsoprinttheloopby
pressingthePrintLoopmodulekey.
SettingPawalarmlimits
Youcansetthelimitalarmsonoroff,andadjusttheiractivationlimitsaccording
toyourneeds.
1.SelectthePatientSpirometryparameterwindow.
2.SelectPawAlarms.
3.Setthealarmlimits:PEEPtot(ORandPACUsoftwarepackages),PEEPiandPEEPe
(othersoftwarepackages),PpeakandMVexp.
NOTE
Ifthefeatureisnotactive,thealarmlimitsaregreyedout.
SelectAlarmOntosetthealarms.
SettingMV/Ventalarmlimits
Youcansetthelimitalarmsonoroff,andadjusttheiractivationlimitsaccording
toyourneeds.
1.SelectthePatientSpirometryparameterwindow.
2.SelectMV/VentAlarms.
3.Setthealarmlimits.
NOTE
Ifthefeatureisnotactive,thealarmlimitsaregreyedout.
SelectAlarmOntosetthealarms.
NOTE
MV/VentAlarmscannotbeselectedoffexceptwhenan
interfaceddeviceisused.
NOTE
Toenableapneaandtechnicalalarmsfromtheventilator,
selecttheVentilatorApneaandTechnicalAlarmscheck
box.
PatientSpirometrymeasurementbasics
PatientSpirometrymeasurementdescription
TheCARESCAPErespiratorymodulesE-sCOVandE-sCAiOV,andthecompactairway
modulesE-COV,E-COVX,E-CAiOV,andE-CAiOVXmeasureairwaypressures,ow,
volumes,complianceandairwayresistancebreathbybreathatthepatient'sairway.
AllparametersaremeasuredthroughD-lite(+)/Pedi-lite(+)owsensorplacedatthe
patient'sairway.
328CARESCAPEModularMonitors2062971-001