Airwaygases
3.Adapterwithsamplinglineconnector
4.Samplinglineconnectoronthewatertrap
Airwaygasesequipmenttopatientconnectionswith
E-miniC,criticalcaresetup
1.E-miniCmodule
2.Gassamplingline
3.Adapterwithsamplinglineconnector
4.Samplinglineconnectoronthewatertrap
Airwaygasesmeasurementsetup
1.Makesurethatthewatertrapcontainerisemptyandproperlyattached.
2.Connectthegassamplinglinetothesamplinglineconnectoronthewatertrap.
3.ConnectthesamplegasoutlettogasscavengingifN
2
Oorvolatileagentsare
used.
4.Turnonthemonitororconnectthemoduletothemonitor.Themonitorperforms
aself-checkforthemodulewhenthemoduleisconnected.Automaticagent
identicationisactivatedinthosemodulesthathavethisfeature.
5.WaituntilthemessageCalibratingdisappears.
6.Connectthesamplinglinetotheairwayadapterortheairwayadaptertothe
ventilatorcircuit.Positiontheadapterwiththesamplingportupwardstominimize
theamountofcondensedwaterpossiblyenteringthesamplingline.
7.Checkthattheairwayadapterconnectionsaretightandthattheadapteris
operatingproperly.
NOTE
Checkthatthesamplelineisconnectedtothewatertrap
beforeconnectingthemoduletothemonitororturningon
themonitor.
NOTE
Tominimizetheamountofdustdrawnintothegassampling
system,alwayskeepthewatertrapconnectedtothemodule.
Whengasmeasurementisnotinuse,youcandisconnectthe
modulefromthemonitortoeliminatetheoperatingsound
ofthegaspump.