Mixedvenousoxygensaturation(SvO₂)
SvO
2
pointstonote
Followthecathetermanufacturer'sinstructionsforinsertingthecatheter.
Forreliablesaturationvalues,thesignalstrengthindicatorshouldbehigherthan
oneasterisk.
DependingontheSvO
2
moduleused,notallSvO
2
measurementsandsettings
areavailabletovieworchange.
SvO₂measurementsetup
SvO
2
equipmenttopatientconnection
1.ModulewithSvO
2
measurementcapability,E-COPSv
2.Opticalmodule
3.Opticalconnector
4.Swan-Ganzthermodilutioncatheter
CheckingtheSvO
2
measurement
1.CheckthattheSvO
2
valueisdisplayedwhenthecatheterisinserted,the
measurementhasbeenstartedandtheWarmingupmessagehasdisappeared.
2.Checkthepositionofthecatheterregularly.
3.CalibrateinvivoandupdatetheHbvalueatleastevery24hours.
SvO
2
measurementonscreen
Signalqualityindicatorfromthecatheter(onlypresentwhentheSvO
2
ismeasured
fromanexternaldevice):
IndicatorExplanation
noasteriskNosignal
*
Poorsignal
**
Fairsignal
***
Goodsignal
276CARESCAPEModularMonitors2062971-001