ECG
Settingalarmlimitsforindividualleads
1.SelecttheSTparameterwindow.
2.SelectAlarms.
3.SelectAlarmOnforanECGleadtoadjustitsalarmlimits.
Ifthealarmislocked,thereisalocksymbolbesidethecheckboxandthe
selectionisnotavailable.
4.SetHighandLowalarmlimitswiththearrows.
Settingrelativealarmlimits
Youcanadjustthehigh/lowalarmlimitssetaroundthecurrentSTvalueforallofthe
individualSTleadsorfortheleadsinaselectedleadgroup.Forexample,whenyou
selectaRelativeAutoLimitsvalueof2mm,thehighlimitissetatthecurrentSTvalue
+2mm,andthelowlimitissetatthecurrentSTvalue-2mm.
1.SelecttheSTparameterwindow.
2.SelectAlarms.
3.SelectRelativeAutoLimits.
4.Settherelativelimitsasneeded:
SetAllLimitswiththearrowsandselectUpdateAll.
SetlimitsforAnteriorleadswiththearrowsandselectUpdateAnterior.
SetlimitsforInferiorleadswiththearrowsandselectUpdateInferior.
SetlimitsforLateralleadswiththearrowsandselectUpdateLateral.
QTdetection
ThisselectionisnotavailablewithcombinationmonitoringorTRAM.
WithMulti-leadQT/QTcAnalysislicenseonly.
TheadministrationofsomedrugtypescanprolongtheQTsegment.MonitoringQT
segmentchangescanhelpidentifyhowthesedrugsareaffectingtheQTsegment.
QT/QTcmeasurementlimitations
NotavailablewithcombinationmonitoringorTRAM.
AtleastonemeasuredV-leadmustbeavailableinorderforthealgorithmto
processQT.
QT/QTcvaluesarecalculatedwith5-leadwire,6-leadwire,or10-leadwireECG
cables.
StartingtheQT/QTcmeasurement
ThisselectionisnotavailablewithcombinationmonitoringorTRAM.
1.SelecttheHRparameterwindow.
2.SelecttheQTtab.
3.SelectOnfromtheQTAnalysislist.
SettingQT/QTcalarms
ThisselectionisnotavailablewithcombinationmonitoringorTRAM.