Messages
Message
Location
ExplanationWhattodo
●T1Calibrationfail/
T2Calibrationfail/
TbloodCalibration
fail
●al.areaCalibrationontheindicated
channelfailed.
●Checkconnections.
●Replacethetransducer.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●T1high/T1low
●T2high/T2low
●T3high/T3low
●T4high/T4low
●Tbloodhigh/Tblood
low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●T2-T1high
●T4-T3high
●Tblood-T1high
●Tblood-T3high
●al.areaMeasureddeltavalueisequalto
oroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●T1temperatureerror/
T2temperatureerror/
T3temperatureerror/
T4temperatureerror
/Tbloodtemperature
error
●al.areaHardwareorcalibrationtest
failureinthemeasurement
device.
●Changethecable.
●Changethemodule.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Temperatureerror
●param.
Hardwareorcalibrationtest
failureinthemeasurement
device.
●Changethecable.
●Changethemodule.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Tempmeasurement
removed
●al.areaAcquisitionmodule,temperature
cable,ortemperatureprobe(s)
havebeenremoved.
●Checkallconnectionsand
reconnectasrequired.
Messagesrelatedtocardiacoutput
measurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●report=reportview
●wavef.=waveformarea
●C.O.menu=cardiacoutputmenu,Measurementtab