Messages
Message
Location
ExplanationWhattodo
T1Calibrationfail/
T2Calibrationfail/
TbloodCalibration
fail
al.areaCalibrationontheindicated
channelfailed.
Checkconnections.
Replacethetransducer.
Iftheproblempersists,
contactauthorizedservice
personnel.
T1high/T1low
T2high/T2low
T3high/T3low
T4high/T4low
Tbloodhigh/Tblood
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
T2-T1high
T4-T3high
Tblood-T1high
Tblood-T3high
al.areaMeasureddeltavalueisequalto
oroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
T1temperatureerror/
T2temperatureerror/
T3temperatureerror/
T4temperatureerror
/Tbloodtemperature
error
al.areaHardwareorcalibrationtest
failureinthemeasurement
device.
Changethecable.
Changethemodule.
Iftheproblempersists,
contactauthorizedservice
personnel.
Temperatureerror
param.
Hardwareorcalibrationtest
failureinthemeasurement
device.
Changethecable.
Changethemodule.
Iftheproblempersists,
contactauthorizedservice
personnel.
Tempmeasurement
removed
al.areaAcquisitionmodule,temperature
cable,ortemperatureprobe(s)
havebeenremoved.
Checkallconnectionsand
reconnectasrequired.
Messagesrelatedtocardiacoutput
measurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
report=reportview
wavef.=waveformarea
C.O.menu=cardiacoutputmenu,Measurementtab
482CARESCAPEModularMonitors2062971-001