ECG
WARNING
—PAUSEDANALYSIS—Certainconditionspausearrhythmia
analysis.Whenpaused,arrhythmiaconditionsarenot
detectedandalarmsassociatedwitharrhythmiasdonot
occur.Conditionscausingpausedarrhythmiaanalysisinclude
arrhythmiaoff,arrhythmiapaused,leadsfail,alarmpause,all
alarmsoff,anddischargedpatient.
WARNING
—FAILURETODETECTLETHALARRHYTHMIA—Alwaysmonitor
ECGforarrhythmiadetectionpurposes.HRcalculatedfrom
pulsatileSpO
2
waveformmaydiffersignicantlyfromECGHR
measuredvalues.UsersshouldbeawarethattheSpO2probe
offandNoSpO2pulsetechnicalalarmsescalatenohigher
thanaMediumpriority.
WARNING
—FAILURETODETECTLETHALARRHYTHMIA—TheSpO
2
parameterpulsatileheartratemeasurementisbasedonthe
opticaldetectionofaperipheralowpulseandthereforemay
notdetectcertainarrhythmias.Thepulseoximetryparameter
shouldnotbeusedasareplacementorsubstituteforECG
basedarrhythmiaanalysis.
WARNING
—ARRHYTHMIAPAUSEDalarm—TheArrhythmiapaused
alarmindicatesthatthesystemisnolongermonitoring
arrhythmiaorheartratefromECG.Ifyouadjustthealarm
prioritylevellowerthanthedefaultvalue,keepthepatient
underclosesurveillance.
Arrhythmiameasurementlimitations
●Sincethearrhythmiadetectionalgorithmsensitivityandspecicityislessthan
100%,sometimestheremaybesomefalsearrhythmiasdetectedandalsosome
truearrhythmiaeventsmaynotbedetected.Thisisespeciallytruewhenthesignal
isnoisy.
●TheECGsizeandQRSwidthsettingsaffectarrhythmiadetectionandheartrate
calculationsensitivity.
●IfQRSamplitudeislow,themonitormightnotbeabletocalculateHRandfalse
Asystolemayoccur.
●Duringthelearningphaseofthealgorithm,arrhythmiadetectionmaynotbe
available.Asaresult,thepatientconditionshouldbecloselymonitoredduring
thelearningphaseandforseveralminutesafterthelearningphasetoallowthe
algorithmtoreachoptimaldetectionperformance.
●NotallECGacquisitiondevicesrecognizethesamesetofarrhythmias.For
example,telemetrymonitoringdevicesandTRAMdonotrecognizeSVTachy,
MultifocalPVCs,andMissingbeat,butthesewillberecognizedbydeviceslike
PDMandPSM.Additionalarrhythmiarecognitionmaybeaddedinfutureversions
ofmonitoringdevices.Becauseremotedeviceslikecentralstationsmaynotbe
awareofallarrhythmiasrecognizedbyacquisitiondevices,themonitoringdevices
renamesomearrhythmiaswhentheyaresenttothenetwork.Monitoringdevices
renamethefollowingarrhythmiaswhensendingthemtothenetwork:
ShownonthemonitorShownatremotedeviceslikecentral
stations
SVTachyTachy
FrequentSVCs
Irregular
176CARESCAPEModularMonitors2062971-001