CO2withCAPNOSTATMainstream,CapnoFlexLF,andDualCO2modules
3.Accessoryport(forsamplelines,nasalcannulasetc.)
4.CapnoFlexLFmodule
5.CapnoFlexLFadapter
6.Exhaustline
PreparingthesetupforCapnoFlexLFmodule
1.MakesuretheCapnoFlexLFadapterisattachedtotheTram-Rac.
2.PlugtheCAPNOSTATMainstreamorDualCO2moduleintotheTram-Rac.
3.PlugtheadaptercableintotheCAPNOSTATMainstreamorDualCO2module.
4.ConnectthesamplelinetotheCapnoFlexLFmodulebutnottothepatient.
5.ConnecttheCapnoFlexLFmoduleintotheCapnoFlexLFadapter.Themessage
Warmingupappearsonthescreen.Waituntilitdisappears.
6.IfthemessageCalibratesensorappears,performcalibration.
7.Blocktheopenendofthesamplelineornasalcannula.ThemessageSample
lineblockedshouldappearwithin60secondsonthemonitordisplaytoensure
thatthemonitorrecognizesblockages.
8.RemovethesamplelinefromtheCapnoFlexLFmoduleandreinsertittoreturn
themoduletonormaloperation.
9.Connectthesamplelinetothepatient’sbreathingcircuit.
CalibratingtheCapnoFlexLFadapter
CapnoFlexLFadaptercalibrationshouldbeperformedwheneverthemessage
Check/calibr.adapterappears.
1.PlacethesamplelineawayfromallsourcesofCO
2
,includingthepatient’sexhaled
breath,yourexhaledbreath,andventilatorexhaustvalves.
2.Withtheadapterconnectedtothesensor,thesensorconnectedtothemonitor,
andthepumpturnedon,selectagasrelatedparameterwindow.
3.SelecttheCalibrationtab.
4.SelectStartCalibrating.
ThemessageCalibratedappearsonthescreenafterasuccessfulcalibration.Ifthe
calibrationfails,themessageCalibrationerrorappearsinstead.Inthiscase,perform
anewcalibration.
EquipmentconnectionwithDualCO2module