CO2withCAPNOSTATMainstream,CapnoFlexLF,andDualCO2modules
WARNING
WithdeactivatedApneaalarm,keepthepatientunderclose
surveillance.
Apneaalarms’deactivationwiththepauseaudiokey
ApneaalarmscanbedeactivatedwiththepauseaudiokeyiftheAllowalarm
deactivationwiththeAudioPausekeyfor:settingApnea(CO2/Imped)isenabledin
theCareUnitSettings.Thissettingispasswordprotected.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
SelectingCO
2
average
YoucanselectatimeintervalforaveragingtheCO
2
.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheCO2Averagelist.
SelectingtheFiO
2
level
ThepresenceofalargeconcentrationofoxygencausestheCO
2
levelappearlower
thantheactualvalue.UsethisoptiontocompensateforthepresenceofO
2
.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheFiO2levellist.
SelectingtheN
2
Olevel
NOTE
Withtheanestheticagentmeasurementlicenseonly.
AvailableforOR,PACUorICUsoftwarepackagesonly.
ThepresenceofN
2
OcausestheCO
2
valuetoappearhigherthantheactualvalue.
UsethisoptiontocompensateforthepresenceofN
2
O.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheN2Olevellist.
Selectingthepumponoroff
Thepumpmustbeonwhenusingthesidestreammethod.Notealsothatablocked
linemaycausethepumptoshutoffautomatically.
PumponoroffwiththeCapnoFlexLFmodule
ThepumpisautomaticallyturnedonwhenasamplelineisinsertedintotheCapnoFlex
LFmodule.Thepumpturnsoffwhenthesamplelineisremoved.
SelectingpumponoroffwiththeDualCO2module
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
316CARESCAPEModularMonitors2062971-001