EasyManua.ls Logo

GE CARESCAPE - Apnea alarms’ deactivation with the pause audio key; Selecting CO2 average; Selecting the FiO2 level; Selecting the N2 O level

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CO2withCAPNOSTATMainstream,CapnoFlexLF,andDualCO2modules
WARNING
WithdeactivatedApneaalarm,keepthepatientunderclose
surveillance.
Apneaalarms’deactivationwiththepauseaudiokey
ApneaalarmscanbedeactivatedwiththepauseaudiokeyiftheAllowalarm
deactivationwiththeAudioPausekeyfor:settingApnea(CO2/Imped)isenabledin
theCareUnitSettings.Thissettingispasswordprotected.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
SelectingCO
2
average
YoucanselectatimeintervalforaveragingtheCO
2
.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheCO2Averagelist.
SelectingtheFiO
2
level
ThepresenceofalargeconcentrationofoxygencausestheCO
2
levelappearlower
thantheactualvalue.UsethisoptiontocompensateforthepresenceofO
2
.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheFiO2levellist.
SelectingtheN
2
Olevel
NOTE
Withtheanestheticagentmeasurementlicenseonly.
AvailableforOR,PACUorICUsoftwarepackagesonly.
ThepresenceofN
2
OcausestheCO
2
valuetoappearhigherthantheactualvalue.
UsethisoptiontocompensateforthepresenceofN
2
O.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheN2Olevellist.
Selectingthepumponoroff
Thepumpmustbeonwhenusingthesidestreammethod.Notealsothatablocked
linemaycausethepumptoshutoffautomatically.
PumponoroffwiththeCapnoFlexLFmodule
ThepumpisautomaticallyturnedonwhenasamplelineisinsertedintotheCapnoFlex
LFmodule.Thepumpturnsoffwhenthesamplelineisremoved.
SelectingpumponoroffwiththeDualCO2module
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
316CARESCAPEModularMonitors2062971-001