ECG
analysisisintendedforuseinhospitals,outpatientclinics,emergencydepartments,
andout-ofhospitalsitessuchasambulancesandpatients’homes.
ACSTooloptionisintendedforadultpatientpopulationwhoaresuspectedclinicallyto
haveacutecoronarysyndrome.
NOTE
Althoughthe12SLanalysisprogramcanbeused
out-of-hospital,CARESCAPEmonitorsareintendedfor
in-hospitaluseonly.
The12SLanalysisprogramisalsoreferredtoasthe12leadECGanalysisprogram.
IntendeduseofACI-TIPI
TheAcuteCardiacIschemia–TimeInsensitivePredictiveInstrument(ACI-TIPI)is
intendedtobeusedinahospitalorclinicenvironmentbycompetenthealth
professionals.TIPIutilizesrecordedECGdataalongwithpatientdemographicand
chestpainstatustoproduceanumericalscorewhichisthepredictedprobabilityof
acutecardiacischemia.Likeanycomputer-assistedECGinterpretationprogram,the
GEMarquetteACI-TIPIevaluationandprobabilityscoreisintendedtosupplement,
notsubstituteforthephysician’sdecisionprocess.Itshouldbeusedinconjunction
withknowledgeofthepatient’shistory,theresultsofaphysicalexamination,the
ECGtracing,andotherclinicalndings.
ACI-TIPIisintendedforadultpatientpopulations.
12leadECGanalysispointstonote
Fora12leadECGanalysis,a12SLECGwithACI-TIPIlicense,10-leadwirecable,and
10–leadleadelectrodeplacementarerequired.
Toobtainthemostaccurate12leadECGanalysis,youshouldenteraccurate
patientdemographics.Thisisespeciallyimportantwhenstoringandcomparing
12leadreportsintheMUSEdatabase.
Fora12leadECGanalysiswiththe12RLfeature,a12RL12leadECGlicenseanda
6–leadwirecable(ora10-leadwirecablewiththeC2/V2,C3/V3,C4/V4,andC6/V6
leadsdisconnected)isrequired.
Fora12leadECGanalysiswiththe12RLfeature,conrmthattheVaandVblead
positionsaresetcorrectlyfora12RLmeasurement.
Forthemostaccurateserialcomparisons,usethesameelectrodecongurationas
usedontheprioranalysisforthepatient.
12RLmeasurementdataisnotsenttotheS/5Network.
Enteringdatafora12leadECGanalysis
WhentheTechIDRequiredissettomandatoryintheCareUnitSettings,youmust
entertheTechnicianIDbeforeyoucanconrmthe12leadsettings.AllCareUnit
Settingsarepasswordprotected.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
1.SelecttheHRparameterwindow.
2.Select12LeadAnalysis.
3.SelectSettings.
2062971-001CARESCAPEModularMonitors167