Messages
Message
Location
ExplanationWhattodo
●Artifact
●param.
Thesamplelinelengthisnot
correct.
●Checkthatthesampleline
lengthis2meters.
●Changesamplelineif
necessary.
●Bypassowhigh
●param.
Themoduleisunableto
synchronizeowandCO
2
dueto
bypassow.
●Adda5mlspacertothe
patientcircuitbetweenthe
Y-pieceandD-lite.
●Shortentheexpirationtimeto
avoidexpiratoryowpause.
●NoVO2,FiN2Ohigh
●param.
ModulehasdetectedN
2
O.●Gasexchangecannot
bemeasuredifN
2
Ois
used.Ifyouwishto
measuregasexchange,
useanotheranesthetic.Ifthe
measurementisnotused,
youmayconsiderremoving
gasexchangenumbersfrom
thescreen.
●NoVO2,FiO2>85%
●param.
MeasuredFiO
2
ismorethan
85%.
●Gasexchangecannotbe
measuredifFiO
2
ismorethan
85%.Ifthepatient’soxygen
valuesareconsistentlyhigh,
youmayconsiderremoving
gasexchangenumbersfrom
thescreen.
●Outofrange
●param.
VO
2
orVCO
2
isbelow0ml/min
orabove999ml/min.
●Checkthatthegassampling
lineandspirometrylinesare
correctlyconnectedtothe
patientairwayandtothegas
module.
●Checkthatthecorrectsensor
type(D-lite/PediLite)has
beenselectedfromthe
monitormenu.
●Servicegasmodule
●Servicegasmodule
-andspecicerror
indication
●al.area,
param.,
wavef.
Themeasuringsensoris
inoperativeorthetemperature
inthemodulehasincreased.
●Contactauthorizedservice
personnel.
MessagesrelatedtoEntropymeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
496CARESCAPEModularMonitors2062971-001