Messages
Message
Location
ExplanationWhattodo
Artifact
param.
Thesamplelinelengthisnot
correct.
Checkthatthesampleline
lengthis2meters.
Changesamplelineif
necessary.
Bypassowhigh
param.
Themoduleisunableto
synchronizeowandCO
2
dueto
bypassow.
Adda5mlspacertothe
patientcircuitbetweenthe
Y-pieceandD-lite.
Shortentheexpirationtimeto
avoidexpiratoryowpause.
NoVO2,FiN2Ohigh
param.
ModulehasdetectedN
2
O.Gasexchangecannot
bemeasuredifN
2
Ois
used.Ifyouwishto
measuregasexchange,
useanotheranesthetic.Ifthe
measurementisnotused,
youmayconsiderremoving
gasexchangenumbersfrom
thescreen.
NoVO2,FiO2>85%
param.
MeasuredFiO
2
ismorethan
85%.
Gasexchangecannotbe
measuredifFiO
2
ismorethan
85%.Ifthepatient’soxygen
valuesareconsistentlyhigh,
youmayconsiderremoving
gasexchangenumbersfrom
thescreen.
Outofrange
param.
VO
2
orVCO
2
isbelow0ml/min
orabove999ml/min.
Checkthatthegassampling
lineandspirometrylinesare
correctlyconnectedtothe
patientairwayandtothegas
module.
Checkthatthecorrectsensor
type(D-lite/PediLite)has
beenselectedfromthe
monitormenu.
Servicegasmodule
Servicegasmodule
-andspecicerror
indication
al.area,
param.,
wavef.
Themeasuringsensoris
inoperativeorthetemperature
inthemodulehasincreased.
Contactauthorizedservice
personnel.
MessagesrelatedtoEntropymeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
496CARESCAPEModularMonitors2062971-001

Table of Contents

Related product manuals