Temperature
Checkingthetemperaturemeasurement
1.Checkthatthetemperaturevalueisdisplayedwhentheprobeisconnectedtoa
singleordualtemperaturecable.
Temperaturemeasurementonthemonitor
screen
Uptofourtemperaturemeasuringsitescanbesimultaneouslymeasuredand
monitored(vesiteswhenmonitoringTblood).Temperaturemonitoringprovides
numericsonly.Nowaveformisgeneratedordisplayed.
NOTE
Amaximumofsixuser-denedmeasuringsitelabelnames
canbecongured.SeetheCARESCAPEModularMonitors
SupplementalInformationManualfordetails.
Thedefaulttemperaturemeasuringsitelabelsareasfollows:
T1,T2=generallabelSkin=skin
T3,T4=generallabel
AirW=airway
Eso=esophageal
Room=room
Naso=nasalMyo=myocardial
Tymp=tympanic
Core=core
Rect=rectalSurf=surface
Axil=axillaryBlad=bladder
Usingthetemperaturemeasurement
Temperaturemappings
Temperaturemeasurementsaremappedtooneofvetemperaturechannelsas
follows:
Temperaturechannel
Measurementsource,B850Measurementsource,B650
orB450
T1
PDM(Masimo),PDM(Nellcor),
TRAM451,TRAM451N5,
TRAM451M,TRAM851,TRAM
851N5,TRAM851M,PSM.
PDM(Masimo),PDM(Nellcor),
PSM.
T2
PDM(Masimo),PDM(Nellcor),
TRAM451,TRAM451N5,
TRAM451M,TRAM851,TRAM
851N5,TRAM851M,PSM.
PDM(Masimo),PDM(Nellcor),
PSM.
T3
E-PTmoduleE-PTmodule
2062971-001CARESCAPEModularMonitors257