PatientSpirometry
PatientSpirometrytroubleshooting
ProblemSolution
ValuesseemerroneousCheckthepatientstatus.
Checkthatyouareusingthecorrectsensortype:
D-lite(+)foradultpatients,Pedi-lite(+)forpediatric
patients.
Checkthesensortypeselection.
CheckthatthePatientSpirometrytubeconnectors
andtheirconnectionsaretightandnotleaking.
Checkthearterialbloodgasvalues.
Checkthatthesamplinglineisnotkinked.
Valuesseemunstable
RemovetheD-lite(+)/Pedi-lite(+)andshakedrops
away.
Checkthattheconnectorsonthe
D-lite(+)/Pedi-lite(+)areintactandthat
connectionsaretight.
StrongvibrationsintheloopCheckthepatientstatus.
Checkthepatientandsystemforwateror
secretions.
Whydothemonitorandtheventilatorshowa
differentcompliancevalue?
Themonitormeasuresdirectlyattheendotracheal
tubeandreectsmoreaccuratecomplianceof
thelung.Hosesandotherequipmentbetween
thepatientandtheventilatorinuencethevalues
measuredintheventilator.