ECG
monitorgeneratesa12leadreport,savesthereportlocally,anddisplaysthereport
onthescreen.Themonitorcanstoreuptofteen12leadreportslocally.
A12leadreportandtheMUSEdatabase
Afterthemonitorgeneratesthe12leadreport,youcansendittotheoptionalMUSE
databaseforfurtheranalysisorstorage,printthereport,ordeletethereport.
TheSendtoMUSEorMUSE+Printarenotselectablewhen:
●TheMUSEdatabaseisnotavailableorwhenthelocalreporthasalreadybeensent
totheMUSEdatabase.
●AtemporarymedicalrecordnumberisusedandCareUnitSettings>Transmit
withtempMRNhasnotbeenenabled.
●The12RLfeaturewasused.TheMUSEdatabasedoesnotsupporta12leadreport
withthe12RLfeature.
●ThemonitorisconnectedtotheS/5Network.
Sendinga12leadECGreporttotheMUSEdatabase
1.SelecttheHRparameterwindow.
2.Select12LeadAnalysis.
3.SelectSendtoMUSE.
Thelocal12leadECGreportisnowsenttotheMUSEdatabase.
4.SelectPrinttoprintthe12leadreport.
YoucanalsoselectMUSE+Printinsteadofthetwoseparatesteps3and4above.
5.SelectDeletetodeletethereportandreturntothereal-timeview.
6.Togenerateanew12leadECGanalysisreport,selectReal-timeViewandrepeat
theprocedureforperforminga12leadECGanalysis.
Viewingorprintingsaved12leadECGreports
Youcanviewandprint12leadreportsthatarestoredatthemonitor(local),orif
available,storedataMUSEdatabase.Thenewestreportsaredisplayedrst.
ToopenareportthatisstoredattheMUSEdatabase,aconnectiontotheCARESCAPE
Networkisrequired.
1.SelecttheHRparameterwindow.
2.Select12LeadAnalysis.
3.SelectSavedReports.
4.Selectthedesired12leadreportfromthelist.
5.Toviewthisreport,selectView.
Areportthatisstoredlocallyatthemonitoropensanddisplaysinthe12Lead
Analysisview.AreportthatisstoredattheMUSEdatabaseopensanddisplaysin
theMUSEReportview.
6.Tosendalocallysaved12leadECGreporttotheMUSEdatabase,selectSend
toMUSE.
YoucanonlysendthereporttotheMUSEdatabaseonce.