ECG
IEC
AAMI/AHA
Electrodeplacement
R(red)RA(white)Rightdeltoidorwrist.
L(yellow)LA(black)Leftdeltoidorwrist.
N(black)
RL(green)
Rightthighorankle.
F(green)
LL(red)Leftthighorankle.
C/C1(white)V/V1(brown)
4
th
intercostalspace,rightborderofthe
sternum.
C2(white/yellow)V2(brown/yellow)
4
th
intercostalspace,leftborderofthe
sternum.
C3(white/green)V3(brown/green)MidwaybetweenC2/V2andC4/V4.
C4(white/brown)V4(brown/blue)
5
th
intercostalspace,mid-clavicularline.
C5(white/black)V5(brown/orange)LeftanterioraxillarylineatC4/V4level.
C6(white/purple)V6(brown/purple)Mid-axillarylineatC4/V4andC5/V5levels.
CheckingtheECGmeasurement
1.Checkthatthewaveformsandparametervaluesaredisplayedwhenthecableis
connectedtothepatient.
AbouttheECGanalogoutputsignal
MaximumdelayoftheECGanalogoutputsignalis30mswiththePDMand15ms
withE-modules.Pacemakerpulseindicationisincludedwhenappropriateanditis
summedintotheECGwaveform.
ECGsynchronizationpulsedelayfromtheR-wavepeakis<35msforthePDMand
E-modules,withtheexceptionofwideQRS(120ms/0.5mV).
Formoreinformationanddetailedspecications,refertothesupplemental
informationmanual.
154CARESCAPEModularMonitors2062971-001