EEGandauditoryevokedpotentials
1.Frontalcortexandassociationareas;latecorticalresponse
1a=P1
1b=N1
1c=P2
1d=N2
2.Medialgeniculateandprimaryauditorycortex;earlycorticalresponse
2a=No
2b=Po
2c=Na
2d=Pa
2e=Nb
3.Acousticnerveandbrainstem;brainstemresponse
AEPtroubleshooting
ProblemSolution
Whyaretherenoclickstobeheardfromthe
earphones/headphones?
Theconnectorisnotinplace.
Checkthattheplugisrmlypushedintothe
appropriateconnectorintheheadbox.
WhatifallormostoftheEPepochsarerejected?ThesignalhastoomuchnoiseorartifactintheEP
measurementband,forinstancecoupled50/60Hz
maynotbeseenintheEEGwaveformbecauseof
ltering.
Checkthattheelectrodeimpedancesarebelow
5kOhm,andthattheimpedancesofasame
channelareclosetoeachother.Ifthisdoesnot
help,trytoremovesourcesofnoise.
392CARESCAPEModularMonitors2062971-001