Messages
Message
Location
ExplanationWhattodo
Headboxoff
param.
Headboxisnotconnected.Connecttheheadbox.
HighEMG
param.
Highfrontalmuscleactivitymay
disturbtheEEGmeasurement.
Calmthepatient.
IdenticalEEG
modules
al.areaTherearetwoormoreEEG
modulesinthesystem.
RemoveallbutoneEEG
module.
Poorelectr.contact
param.
Electrodeimpedanceisover5
kOhm.
Presstheelectrodeto
improveconnection.
Changetheelectrode.
Checktheimpedance.
MessagesrelatedtoBISmeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
report=reportview
wavef.=waveformarea
Message
Location
ExplanationWhattodo
Applysensor
param.
Thesensorconnectiontothe
patientmaybeloose.
PresstheBISsensor
electrodestoimprove
connection.
Artifact
param.,
wavef.
Signalscontainnoiseorartifact.Checksensorcontact.
Removesourcesofexcessive
noise.
Automaticcheckoff
param.
Automaticsensorcheckhas
beenturnedoff.
Ifrequired,activatethe
automaticcheck.
BIScableoffal.areaCableisdisconnectedfromthe
module.
Connectthecable.
BISDSCerror
al.areaTheBISxisnotcommunicating
oroperatingproperly.
TesttheBISx.
Iftheproblempersists,
contactauthorizedservice
personnel.
BIShigh/BISlow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Checkthedosageof
anesthetics.
Adjustalarmlimitsif
necessary.
2062971-001CARESCAPEModularMonitors501