Cleaningandcare
AboutPDMbatterycharging
B850:ThePDMbatteryischargedwheneverthePDM,withitsbatteryinstalled,is
connectedtoamonitorthatisconnectedtoanAC-derivedpowersource.Thebattery
capacitygauge(labeledwithtextPDM)indicatesthebattery’schargelevel.
B650andB450:ThePDMbatteryisnormallychargedwheneverthePDM,withits
batteryinstalled,isconnectedtothemonitorunlessthemonitorbatteryisbeing
chargedorprintingtoarecorderisactive.
YoucanalsochargePDMonanexternalcharger.Refertothecharger’sinstructions
foruse.
ReplacingthePDMbattery
WARNING
—EXPLOSIONORFIRE—Usingnon-recommendedbatteries
couldresultininjury/burnstothepatientsorusers.Onlyuse
batteriesrecommendedormanufacturedbyGE.Thewarranty
canbevoidedifnon-recommendedbatteriesareused.
WARNING
—PHYSICALINJURY—DonotinstallthePDMaboveapatient.
Makesurethebatteryiscompletelyinsertedandthebattery
dooriscompletelyclosed.Fallingbatteriescouldseriouslyor
fatallyinjureneonatalorothervulnerablepatients.
WARNING
—PHYSICALINJURY—DonotinstallthePDMabovea
patient.Leaksfromthebatterycellscanoccurunderextreme
conditions.Theliquidiscaustictotheeyesandskin.Ifthe
liquidcomesincontactwitheyesorskin,ushwithclean
waterandseekmedicalattention.
1.Openthebatterydoorbygentlypullingonthebatterydoorpulltab.
2.PullthebatterytrayoutofthePDMusingthebatterytraystrapandremovethe
batteryfromthebatterytray.
3.Insertthenewbatterywiththetestbuttonfacingupandthearrowpointinginto
thePDM.
4.Pressthebatterydoorcloseduntilitsealsthebatterycompartment.
5.ConnectthePDMtothemonitor.
6.ConrmthePDMbatterycapacitygaugedisplaysonthemonitor.
Batteryrecycling
Whenabatterynolongerholdsacharge,itshouldbereplaced.Removetheold
batteryandfollowyourlocalrecyclingguidelines.
WARNING
Donotincinerateabatteryorstoreathightemperatures.
Seriousinjuryordeathcouldresult.
Abouttheinternallithiumbattery
Themonitorcontainsalithiumbattery.Thisbatteryretainsthecorrecttimeanddate
onthemonitor.
460CARESCAPEModularMonitors2062971-001