Airwaygases
WARNING
WithdeactivatedApneaalarm,keepthepatientunderclose
surveillance.
Apneaalarms’deactivationwiththepauseaudiokey
ApneaalarmscanbedeactivatedwiththepauseaudiokeyiftheAllowalarm
deactivationwiththeAudioPausekeyfor:settingApnea(CO2/Imped)isenabledin
theCareUnitSettings.Thissettingispasswordprotected.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
SelectingwhattoshowwithEtCO
2
Youcanselectwhichothergasmeasurementvalueappearsintheparameterwindow
withtheEtCO
2
.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheShowwithEtCO2list.
SelectingtheFiO
2
level
NOTE
E-miniCandOR,PACU,ED,andICUsoftwarepackagesonly.
NOTE
FiO
2
andN
2
Ocompensationsmustbeselectedmanually
whenE-miniCisused.
ThepresenceofalargeconcentrationofoxygencausestheCO
2
levelappearlower
thantheactualvalue.UsethisoptiontocompensateforthepresenceofO
2
.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheFiO2levellist.
SelectingtheN
2
Olevel
NOTE
E-miniCandtheanestheticagentmeasurementlicenseonly.
AvailableforOR,PACU,orICUsoftwarepackages.
NOTE
FiO
2
andN
2
Ocompensationsmustbeselectedmanually
whenE-miniCisused.
ThepresenceofN
2
OcausestheCO
2
valuetoappearhigherthantheactualvalue.
UsethisoptiontocompensateforthepresenceofN
2
O.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheN2Olevellist.