Non-invasivebloodpressure
NIBPtroubleshooting
ProblemSolution
NIBPmeasurementdoesnotworkor
thevaluesseemunstable.
Checkthatthecufftubingisnotbent,stretched,compressed,
orloose.
Checkthecuffpositionandcufftubeconnection.
Preventmotionartifact.
UseNIBPcuffsofcorrectsize.
Whydoesthemeanvaluedisplaywhile
theassociatedsystolicanddiastolic
valuesdisplayas---?
Assessthepatientandperformavisualinspectionofthe
equipmenttoensuresystemintegrity.
Thefollowingconditionsmaycausethemeanvaluetodisplayin
theNIBPparameterwindowwhiletheassociatedsystolicand
diastolicvaluesdisplayas---:
Verylowsystolicanddiastolicamplitudeuctuations(e.g.,
patientinshock).
Verysmalldifferencebetweenthemeanandsystolicpressure
orthemeananddiastolicpressure.
Lossofsystemintegrity(e.g.,looseconnectionsorwornparts)
Whyisthemonitorre-inatingthecuff
automatically?
Thecufftargetpressuremustbehigherthanthepatient’ssystolic
pressuretoobtainanaccuratesystolicanddiastolicmeasurement.
Ifasystolicbloodpressurecannotbefound,themonitorsearches
forasystolicreadingbyre-inatingthecuffatahigherpressure.
ThissystolicsearchmayoccuronceperNIBPdeterminationcycle.
Duringasystolicsearch,themaximumcuffinationpressure
willnotexceedthenormalpressurerangeofthecuff.Formore
information,refertothetechnicalspecications.
ForPDMandPSM,themonitormaybetakingacontrol
measurement.IfthemeasuredNIBPvalueexceedsthealarm
limits,asinglelowpriorityalarmsoundsandanewmeasurement
isautomaticallytaken.Ifthenewvalue(thecontrolmeasurement)
alsoexceedsthealarmlimitsthealarmpriorityescalatesto
medium.InManualmodeandSTATmodethereareatleast
foursecondsbetweentherstmeasurementandthecontrol
measurementforAdultandChildcuffs,eightsecondsforInfant
cuffs.InAutomodethereareatleast30secondsbetweentherst
measurementandthecontrolmeasurement.
ForPSM,ifWeakpulsationorArtifactsisdetected,themodule
repeatsthemeasurementuptothreetimes.Assessthepatient
andperformavisualinspectionoftheequipmenttoensuresystem
integrity.
ForPSM,ifcuffocclusionisdetectedduringinationthemodule
mightrepeatthemeasurement.Performavisualinspectionofthe
equipmenttoensuresystemintegrity.
2062971-001CARESCAPEModularMonitors233