Cleaningandcare
Cleaningandcaresafetyprecautions
Cleaningandcarewarnings
WARNING
Beforecleaningordisinfecting,disconnectthemonitorfrom
thepowersupply.
WARNING
Regularpreventivemaintenanceshouldbecarried
outannually.Failuretoimplementtherecommended
maintenanceschedulemaycauseequipmentfailureand
possiblehealthhazards.
WARNING
Theusermayonlyperformmaintenanceprocedures
specicallydescribedinthismanual.
WARNING
Non-medicalequipmentdoesnotprovidethesamelevelof
protectionagainstelectricalshock.Donottouchthepatient
andanypartofnon-medicalequipmentatthesametime.
Someexamplesofnon-medicalequipmentarelaserprinters
andnon-medicalcomputers.
WARNING
Avoiduseofcleaners,materialsorchemicalsthatmay
damagedevicesurfaces,labels,orcauseequipmentfailures.
WARNING
Topreventliquidsfromenteringthemonitorordisplaycasing,
donottiltthemonitorordisplaymorethan+/-15degrees.
WARNING
Ifliquidhasaccidentallyenteredthesystemoritsparts,
disconnectthepowercordfromthepowersupplyandhave
theequipmentservicedbyauthorizedservicepersonnel.
WARNING
UseonlywashablekeyboardwithatleastIPX1protection
againstingressofwater.
WARNING
Cleanupanddisposalofbrokendisplaymonitorsmustbe
incompliancewiththesafetyandwastecontrolguidelines
regulatingthisproduct.
WARNING
Neverimmerseanypartofthedevice,cables,orleadwiresin
liquidsorallowliquidtoentertheinteriorofthedevice.
2062971-001CARESCAPEModularMonitors453