Interfacingwithperipheraldevices
■ThemonitordisplaysPpeak,Pplat,Pmean,PEEPtot,PEEPi,PEEP,TVexp,TVinsp,
MVexp,MVspont,Compl,Raw,I:Eratio,andstaticComplmeasurementdata.
■Ifthereisacompactairwaymodulemeasuringspirometryinthemonitor,itis
usedasthemeasurementsourceinsteadoftheinterfaceddevice.
●Gasdatafromventilator,anesthesiamachines,andgasanalyzers
■ThemonitordisplaysCO
2
,RespRate,O
2
,N
2
O,Halothane,Desurane,Enurane,
IsouraneandDesuranemeasurementdata.
■Ifthereisacompactairwaymoduleinthemonitor,itisusedasthe
measurementsourceinsteadoftheinterfaceddevice.
●Laboratorydata
■ThemonitorprovidesLaboratoryData>ViewandEnterDatamenus.
●Continuouscardiacoutput
■TheinterfacedcontinuouscardiacoutputmeasurementprovidesC.O.,blood
temperature,CCO,SVV,GEDI,ELWIandSVRmeasurementsdependingon
theinterfaceddevice.TheExtindicatesthedataisfromanexternalsource.
Thefollowingconditionsapplywiththeinterfacedcontinuouscardiacoutput
measurement:
♦TheC.I.valueforinterfaceddeviceiscomputedusingthefollowingequation
C.I.=C.O./BSA.
♦TheC.I.valueforinterfaceddeviceisinvalidifBSAfromthemonitorisnot
available.
♦TheCCIvalueforinterfaceddeviceiscomputedusingthefollowingequation
CCI=CCO/BSA.
♦TheCCIvalueforinterfaceddeviceisinvalidifBSAfromthemonitorisnot
available.
♦TheSVRvaluefortheinterfaceddeviceshallbecomputedusingthe
followingequationSVR=((MAP-CVP)/CCO)*79.92.
♦TheSVRvaluefortheinterfaceddeviceshallbeinvalidifMAP,CVPorCCO
fromthemonitorarenotavailable.
♦TheSVRIvalueforinterfaceddeviceshallbecomputedusingthefollowing
equationSVRI=((MAP-CVP)/CCI)*79.92.
♦TheSVRIvaluefortheinterfaceddeviceshallbeinvalidifMAP,CVPorCCI
fromthemonitorarenotavailable.
♦Tbloodvalueisdisplayed.TheTblood-T1temperaturealarmlimitis
adjustable.
●Ventilator/anesthesiamachinesettingsandtechnicalalarms
■Thesettingsaredisplayedinnumerictrends.
■Technicalalarmsandsettingsaredisplayedandtrendedevenwhenthereisa
gasmoduleconnectedtothemonitor.
452CARESCAPEModularMonitors2062971-001