Interfacingwithperipheraldevices
ThemonitordisplaysPpeak,Pplat,Pmean,PEEPtot,PEEPi,PEEP,TVexp,TVinsp,
MVexp,MVspont,Compl,Raw,I:Eratio,andstaticComplmeasurementdata.
Ifthereisacompactairwaymodulemeasuringspirometryinthemonitor,itis
usedasthemeasurementsourceinsteadoftheinterfaceddevice.
Gasdatafromventilator,anesthesiamachines,andgasanalyzers
ThemonitordisplaysCO
2
,RespRate,O
2
,N
2
O,Halothane,Desurane,Enurane,
IsouraneandDesuranemeasurementdata.
Ifthereisacompactairwaymoduleinthemonitor,itisusedasthe
measurementsourceinsteadoftheinterfaceddevice.
Laboratorydata
ThemonitorprovidesLaboratoryData>ViewandEnterDatamenus.
Continuouscardiacoutput
TheinterfacedcontinuouscardiacoutputmeasurementprovidesC.O.,blood
temperature,CCO,SVV,GEDI,ELWIandSVRmeasurementsdependingon
theinterfaceddevice.TheExtindicatesthedataisfromanexternalsource.
Thefollowingconditionsapplywiththeinterfacedcontinuouscardiacoutput
measurement:
TheC.I.valueforinterfaceddeviceiscomputedusingthefollowingequation
C.I.=C.O./BSA.
TheC.I.valueforinterfaceddeviceisinvalidifBSAfromthemonitorisnot
available.
TheCCIvalueforinterfaceddeviceiscomputedusingthefollowingequation
CCI=CCO/BSA.
TheCCIvalueforinterfaceddeviceisinvalidifBSAfromthemonitorisnot
available.
TheSVRvaluefortheinterfaceddeviceshallbecomputedusingthe
followingequationSVR=((MAP-CVP)/CCO)*79.92.
TheSVRvaluefortheinterfaceddeviceshallbeinvalidifMAP,CVPorCCO
fromthemonitorarenotavailable.
TheSVRIvalueforinterfaceddeviceshallbecomputedusingthefollowing
equationSVRI=((MAP-CVP)/CCI)*79.92.
TheSVRIvaluefortheinterfaceddeviceshallbeinvalidifMAP,CVPorCCI
fromthemonitorarenotavailable.
Tbloodvalueisdisplayed.TheTblood-T1temperaturealarmlimitis
adjustable.
Ventilator/anesthesiamachinesettingsandtechnicalalarms
Thesettingsaredisplayedinnumerictrends.
Technicalalarmsandsettingsaredisplayedandtrendedevenwhenthereisa
gasmoduleconnectedtothemonitor.
452CARESCAPEModularMonitors2062971-001