GE CARESCAPE - C.O. measurement setup; C.O. equipment to patient connection with an in-line probe

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cardiacoutput
TheC.O.connectorcablesaremodule-specicandcanonlybeusedwiththe
appropriateC.O.module.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
Thepatient'sheightandweightvaluesarerequiredfordeterminingCardiacIndex
(C.I.).
E-modulesonly:Predenedcathetersarealreadyconguredforarightventricular
ejectionfraction(REF)measurement,butuser-denedcathetersmustbe
conguredtosupportit.Cathetersthatshowupintheselectionlistcanbeadded
ordeletedthroughCareUnitSettings>Parameters>Catheters,andthese
settingsarepasswordprotected.
Bloodtemperaturemeasurement,Tblood,isprovidedfromanysupported
acquisitionmodulethatprovidesC.O.measurements.TheUnityNetworkInterface
Device(ID)canonlybeusedifnootheracquisitionmoduleprovidesC.O.
Dependingonthemoduleused,notallcardiacoutputmeasurementsandsettings
areavailabletovieworchange.
C.O.measurementsetup
C.O.equipmenttopatientconnectionwithanin-lineprobe
1.ModulewithC.O.measurementcapability
2.Cardiacoutputcable
3.In-lineinjectateprobe
4.Injectatesyringe
5.Injectatesolution
6.CVPlinetoIPtransduceroruidinfuser
7.Proximalinjectateport
8.PAdistalport
9.Opticalmoduleconnector(usedforSvO
2
measurement)
262CARESCAPEModularMonitors2062971-001