Cardiacoutput
●TheC.O.connectorcablesaremodule-specicandcanonlybeusedwiththe
appropriateC.O.module.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
●Thepatient'sheightandweightvaluesarerequiredfordeterminingCardiacIndex
(C.I.).
●E-modulesonly:Predenedcathetersarealreadyconguredforarightventricular
ejectionfraction(REF)measurement,butuser-denedcathetersmustbe
conguredtosupportit.Cathetersthatshowupintheselectionlistcanbeadded
ordeletedthroughCareUnitSettings>Parameters>Catheters,andthese
settingsarepasswordprotected.
●Bloodtemperaturemeasurement,Tblood,isprovidedfromanysupported
acquisitionmodulethatprovidesC.O.measurements.TheUnityNetworkInterface
Device(ID)canonlybeusedifnootheracquisitionmoduleprovidesC.O.
●Dependingonthemoduleused,notallcardiacoutputmeasurementsandsettings
areavailabletovieworchange.
C.O.measurementsetup
C.O.equipmenttopatientconnectionwithanin-lineprobe
1.ModulewithC.O.measurementcapability
2.Cardiacoutputcable
3.In-lineinjectateprobe
4.Injectatesyringe
5.Injectatesolution
6.CVPlinetoIPtransduceroruidinfuser
7.Proximalinjectateport
8.PAdistalport
9.Opticalmoduleconnector(usedforSvO
2
measurement)
262CARESCAPEModularMonitors2062971-001