Messages
Message
Location
ExplanationWhattodo
●Zerofailure
●param.
Zeroinghasfailed.●Checkthepatient’spressure
byalternativemeans.
●Replacethemodule.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Zeroing
●param.
Zeroingisinprogress.
●Waituntilthezeroingis
completed.
●ZeroOK
●param.
Zeroingwassuccessful.●Noactionrequired.
Messagesrelatedtoinvasivepressures
measurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●report=reportview
●wavef.=waveformarea
Message
Location
ExplanationWhattodo
●>320mmHgor>43
kPa
●param.
Measurementisoverrange,or
thesensorisfaulty.
Ifyoupre-zeroalinewiththe
stopcockclosed,itcreatesa
highuidbagpressureand
triggersthismessage.Inthis
case,youcanacknowledgethe
alarmwiththepauseaudiokey.
●Checkthepatient’spressure
byalternativemeans.
●Checkthecableand
connections.
●Rezerothetransducer.
●Replacethesensor.
●Replacethetransducer.
●Replacethemodule.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●<-40mmHgor<-5
kPa
●param.
Measurementisunderrange,or
thesensorisfaulty.
●Checkthepatient’spressure
byalternativemeans.
●Checkthecableand
connections.
●Rezerothetransducer.
●Replacethesensor.
●Replacethetransducer.
2062971-001CARESCAPEModularMonitors475