Non-invasivebloodpressure
Ensurethatthepatientisrelaxedanddoesnottalkduringthemeasurement.
NIBPmeasurementsetup
NIBPequipmenttopatientconnection
1.ModulewithNIBPmeasurementcapability
2.Cuffhose
3.Cuffofcorrectsize
4.Brachialarteryarrow(printedoncuff)
5.Cuffindexline(printedoncuff)
NIBPmodulekeys
TherearetwoNIBPrelatedkeysonthePSM:
AutoOn/Off
Startsandstopsautomaticmeasurementsattimed
intervals.
StartCancelStartsasinglemeasurement,andcancelsany
measurementinprogress.
NOTE
YoucanalsoselectNIBPStart/NIBPCancelorNIBPAuto
Start/NIBPAutoStopfromthemonitor’smainmenu.
PreparingtheNIBPpatientconnection
1.SelectanappropriateNIBPcuffsizeforthepatient.
2.ConnecttheNIBPcuffhosetothemodule'sNIBPconnector.
3.PositiontheNIBPcuffonthepatient:
Placethecuffarrowoverthebrachialartery(orwhateverarteryisbeingused).
Makesurethatthecuffindexlinefallswithintherangemarkingsonthecuff.
Wrapthecuffaroundthelimb.
4.MakesurethattheNIBPcufftubesarenotkinked,compressed,orstretched.
5.VerifyorselectthecorrectInit.PressureorCuffSizefromtheNIBPSetupmenu.
CheckingtheNIBPmeasurement
1.Checkthatthepressurevaluesaredisplayed.
222CARESCAPEModularMonitors2062971-001