EasyManua.ls Logo

GE CARESCAPE - SpO₂ measurement setup; SpO2 equipment to patient connection; Preparing the SpO2 connection; Checking the SpO2 measurement

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pulseoximetry
SpO₂measurementsetup
SpO
2
equipmenttopatientconnection
1.AcquisitionmodulewithSpO
2
measurementcapability
2.Interconnectcable
3.Reusablesensors
4.Disposablesensors
PreparingtheSpO
2
connection
1.ConnecttheSpO
2
module(s)tothemonitor.
2.Connecttheadaptercable(s)totheSpO
2
moduleconnector(s).
3.Cleanthesurfaceofreusablesensors.
4.Preparetheapplicationsite(s).
5.Removenailpolishandearrings.
6.Followthesensormanufacturer'sinstructionstopositionthesensor(s).
7.Attachthesensor(s)tothepatient.
8.Stabilizethesensorcable(s)tominimizesensormovement.
CheckingtheSpO
2
measurement
1.Checkthattheredlightislitinthesensor.
2.Checkthatthewaveformsandparametervaluesaredisplayedwhenthesensoris
connectedtothepatient.
SpO
2
functionaltesters
Somemodelsofcommerciallyavailablebench-topfunctionaltestersandpatient
simulatorscanbeusedtoverifyproperfunctionalityofpulseoximetersensors,cables
andmonitors.Whilesuchdevicesmaybeusefulforverifyingthatthepulseoximetry
2062971-001CARESCAPEModularMonitors207