CO2withCAPNOSTATMainstream,CapnoFlexLF,andDualCO2modules
3.SelectOnorOfffromthePumplist.
Selectingapneaalarmlimit
Theapneaalarmcanbeenabledtoselectatimedelaybeforeanapneaalarmis
generated.ThedelayisdenedwiththeApneaLimitSecondssetting.Ifanythingelse
thanthedefault(20seconds)isselected,theapneaalarmdisplaysintheparameter
windowwiththenumberofsecondsselectedforthedelay.
1.Selectagasrelatedparameterwindow.
2.SelecttheCO2tab>Setup.
3.SelectanoptionfromtheApneaLimitSecondslist.
Preventingoperatingroompollutionwith
CAPNOSTATMainstream,DualCO2,and
CapnoFlexLFmodules
WhenN
2
Oandvolatileanestheticsareused,preventoperatingroompollutionby
connectingthemonitor’ssamplegasoutlettothescavengingsystemorbyreturning
thesamplegastothepatientcircuit.
Scavengingtoscavengingsystems
Thegasexhaustmaybescavengedusingthescavengingadapterpackage.
1.Removetheexhaustadapterandtubefromthepackage.
2.Attachtheconnectorendoftheexhausttubetothemoduleoutletmarked
EXHAUST.
3.Installtheexhaustadapterintothegasscavengingsystemoftheanesthesia
deliverysystem,followingtheanesthesiamachinemanufacturer’srecommended
procedure.
4.Drapetheexhausttubesothatitdoesnotinterferewiththeworkarea.
StoppingtheCO
2
measurement
1.Removetheaddedadaptersfromthepatient’sbreathingcircuitandgas
scavenging.
2.Checkthepatient’sbreathingcircuit.
3.Removethegasmodulefromthemonitorwhenitisnotused.
BasicsofCO₂measurement
NormalCO
2
waveform
TheCO
2
waveformisreferredtoascapnogramanditreectsthedifferentstagesin
breathing.Thecapnogramofahealthypatientundercontrolledventilationhasa
normalshape.ChangesintheCO
2
waveformmayindicatecompromisedpatient
respiratoryand/orcirculatoryfunctionorimpropermechanicalventilatorfunctionality.
2062971-001CARESCAPEModularMonitors317