EEGandauditoryevokedpotentials
UsingtheEEGmeasurement
StartingtheEEGmeasurement
EEGmeasurementstartsautomatically.
1.Observetheresultsoftheelectrodecheck.
2.Reconnecttheelectrodesiftheimpedanceistoohigh.
SelectingyourownEEGmontage
Ifyouareusingyourownmontage,youwillhavetoselectitfromthelist.
Preconguredmontageswillbeautomaticallyrecognized.
1.SelecttheEEGparameterwindow.
2.SelectEEGtab>Montage.
3.SelectthemontagefromtheSelectMontagelist.
NOTE
Sincethemonitorsetsleadpositionsautomaticallyaccording
totheidenticationpin,makesurethatyouuseprecongured
leadpositions.Ifyouchangethem,alwaysupdatethe
montagesettingsaccordingtotheinstructions.
SelectingtheEEGscale
ThissettingdenesthescaleforEEGparameterwindows,waveformandspectral
displays
1.SelecttheEEGparameterwindow.
2.SelectEEGtab>Setup.
3.SelectavaluefromtheScaleµVlist.
SelectingtheEEGsweepspeed
ThissettingdeterminesthedrawingspeedfortheEEGwaveform.
1.SelecttheEEGparameterwindow.
2.SelectEEGtab>Setup.
3.SelectavaluefromtheEEGSweepSpeedlist.
Thesmallerthevalue,theslowerthesweepspeed.
NOTE
ThissettingisavailableinBIS,EEGandEntropysetups.
Regardlessofwhereyouchangeit,itwillaffectallthree
parameters.
SelectingEEGnumericparameters
EEGnumerictrendsareavailableintheED,ICU,andNICUsoftwarepackages.
1.SelecttheEEGparameterwindow.
2.SelectEEGtab>Setup.
3.SelectaparameterfromtheNumeric1list.
2062971-001CARESCAPEModularMonitors379