Safety
Electromagneticcompatibility
EMCwarnings
WARNING
Otherequipmentmayinterferewiththesystem,evenifthat
otherequipmentcomplieswithCISPRemissionrequirements.
WARNING
Useonlyapprovedaccessories,includingmounts,and
debrillator-proofcablesandinvasivepressuretransducers.
Foralistofapprovedaccessories,seetheCARESCAPE
ModularMonitorsSupplementalInformationManual.Other
cables,transducersandaccessoriesmaycauseasafety
hazard,damagetheequipmentorsystem,resultinincreased
emissionsordecreasedimmunityoftheequipmentorsystem
orinterferewiththemeasurement.
WARNING
PinsofconnectorsidentiedwiththeESDwarningsymbol
shouldnotbetouched.Connectionsshouldnotbemade
totheseconnectorsunlesselectrostaticdischarge(ESD)
precautionsareused.
WARNING
Donotusethemonitorinhighelectromagneticelds(for
example,duringmagneticresonanceimaging).
EMCcautions
CAUTION
UseofknownRFsources,suchascell/portablephones,
orotherradiofrequency(RF)emittingequipmentnearthe
systemmaycauseunexpectedoradverseoperationof
thisdevice/system.Consultqualiedpersonnelregarding
device/systemconguration.
CAUTION
Thedevice/systemshouldnotbeusedadjacentto,orstacked
with,otherequipment.Consultqualiedpersonnelregarding
device/systemconguration.
CAUTION
—EMC—Magneticandelectricaleldsarecapableof
interferingwiththeproperperformanceofthedevice.For
thisreasonmakesurethatallexternaldevicesoperated
inthevicinityofthemonitorcomplywiththerelevantEMC
requirements.X-rayequipmentorMRIdevicesareapossible
sourceofinterferenceastheymayemithigherlevelsof
electromagneticradiation.Changesormodicationstothis
device/systemnotexpresslyapprovedbyGEmaycauseEMC
issueswiththisorotherequipment.Thisdevice/systemis
designedandtestedtocomplywithapplicablestandards
andregulationsregardingEMCandneedstobeinstalled
andputintoserviceaccordingtotheEMCinformation
statedasfollows:Thisdevice/systemissuitableforusein
allestablishmentsotherthandomesticandthosedirectly
connectedtothepubliclow-voltagepowersupplynetwork
thatsuppliesbuildingsusedfordomesticpurposes.Mains
powershouldbethatofatypicalcommercialorhospital
environment.DeviceiscomplianttoClassA.
2062971-001CARESCAPEModularMonitors51